Specific Absorption Rate (SAR) Limit Pop-up Created October 11, 2021 Author Arina Category MRI Troubleshoot What is SAR? Will update later. What We Can Do When SAR Limit Window is Pop-up? Do not simply click Adjust Repetition Time (TR) or Echo Time (TE)If Reduce Flip Angle option is available, you can use this option. If not, please use this step:Click Open ProtocolClick Sequence — Part 2Adjust the RF pulse type & Gradient mode to normal/lowLow SAR Was this article helpful? Yes No